TRUSTCHECK BPM 2.0
The device uses the oscillometric method of blood pressure measurement. It means the device detects the movement of your blood through your brachial artery and converts your blood pressure into a digital reading. An oscillometric monitor does not need a stethoscope, hence the device is simple to use.
▲ The device could measure the systolic pressure, diastolic pressure, and pulse. The components included in the box are the device, cuff, and printed instruction manual. However, batteries are optional. The device is intended to be used by adults only.
▲ Intelligent inflation reduces the uncomfortable feeling caused by incorrect inflation, shortens the measurement time, and prolongs the cuff's usage lifetime.
The device automatically stores 2x99 sets of measurement values and each measurement result will be displayed on the screen. The stored data can be read conveniently by pressing the memory button. This device has a blood classification index which helps to easily check your blood pressure.
Specification:
Description |
Automatic upper arm blood pressure monitor |
|
Display |
LCD digital display |
|
Measuring principle |
Oscillometric method |
|
Measuring localization |
upper arm |
|
Measurement range |
Pressure |
0~299 mmHg (0~39.9kPa) |
Pulse |
40~180 pulses/min |
|
Accuracy |
Pressure |
±3mmHg (±0.4kPa) |
Pulse |
±5% of reading |
|
LCD indication |
Pressure |
3 digits display of mmHg |
Pulse |
3 digits display |
|
Symbol |
Memory/Average/IHB/Heartbeat/Low Battery,etc. |
|
Cuff pressure |
0-299 mmHg |
|
Memory function |
2x99 sets memory of measurement values |
|
Power source |
4pcs AA alkaline battery DC. 6V or USB 5V power guppy |
|
Automatic power off |
In 2 minutes |
|
Main unit weight |
Approx. 235g (batteries not included) |
|
Main unit size |
L129mm x W 100mm x H68mm |
|
Main unit lifetime |
10.000 times under normal use |
|
Battery life |
Could be used for 300 times for normal condition |
|
Accessories |
Cuff. instruction manual |
|
Operating environment |
Temperature |
10~40degree C |
Humidity |
15%~90%RH |
|
Air pressure |
80kPa--105kPa |
|
Storage environment |
Temperature -20 Degree C ~ 5 Degree C. Humidity : 10% ~ 95% avoid crash. sun burn or rain during transportation |
|
Blood Pressure indication Range |
SYS: 60~260mmHg |
|
DIA: 30~200mmHg |
TCBPM2.0